The FPK Downflow Bench range is ideal for many dry dust and fume applications. FPK benches are designed for comfortable seated work with knee-space .
Applications.
- Atmospheric dusts and high volume fumes
- Flour or similar powder weighing and processing
- Glass fibre Reinforced Plastic (GRP) trimming and finishing
- Dry stone cutting, ceramics
- Airline cleaning of electrical components and PCs
- Tool cleaning.
- Hazardous Vapours
- And More
Features.
- Multiple sizes to fit different working areas
- Multiple filter choices for many applications
- Designed for seated work.
- Filter monitor gauge
- Heavy duty
- Replaceable filters
- 230v Plug and Play
Options
- Speed Control
- Draft Skirts
- Rubber Surface Mats
- 3 Sided Surrounds
- Enclosures (3 sides and top)
- Hours Run Meters
- Custom Builds
The FPK Range is constructed from fully welded 2mm, powder coated steel panels. The surface is constructed from 3mm perforated steel plate with a surface loading weight of 100kgs. The FPK Range is an ideal extraction solution for medium to heavier and continuous industrial applications with added knee space for operators to sit comfortably at the bench.
Simply put the FPK Downdraft Bench in place, plug it in and away you go in true Plug And Play Style. Easy to set up and easy to use.
Filters are positively clamped against soft rubber seals; filters can be quickly accessed via the front flap. Replacement is via lowering of the hinged front cover. Filter changes can be accomplished in minutes.
The FPK Range complies with HSE and COSHH requirements for controlling airborne contaminants in the work place.
Model | Dimensions (mm) | Ventilated Area (mm) |
V400-FPK066084 | 674 x 660 x 840 | 610 x 356 |
V400-FPK126084 | 1274 x 660 x 840 | 1210 x 356 |
V400-FPK069084 | 674 x 960 x 840 | 610 x 656 |
V400-FPK129084 | 1274 x 960 x 840 | 1210 x 656 |
V400-FPK189084 | 1874 x 960 x 840 | 1810 x 656 |
Filter Type | Pre Filter | Main Filter | Applications |
A – Non Sparking Dusts | No | G4/100: 95% @ 5 Microns | Course dusts. Plastics, GRP, Atmospheric Dusts or Similar |
F – Fine Dusts | G4/50: 85% @ 4 Microns. | F7: 95% @ 1 Micron | Carbon Fibre Sanding, Powder Sampling, Other Composites, Softwoods, Tool Cleaning (non RCS). General Dust |
B – Very Fine Dust | G4/50: 85% @ 4 Microns | F9: 95% @ 0.5 Microns | Fingerprinting, Rapid Prototyping, Conservation and Restoration, Printer & Toner, Some Chemical Dusts |
L – HEPA | G4/50: 85% @ 4 Microns | H13: 99.99% @ 0.3 Microns | Pharmaceuticals, Hardwoods & MDF, RCS Materials, Hazardous Dust, Powder Sampling, Laboratory |
C – Dust and Light Sparks | Labyrinth & Media Pad: 98% @ 10 Microns | M6 Flame Retardant: 95% @ 0.5 Microns | Fettling, Metal Sanding, Deburring, Weld Dressing |
E – Low Volume Fume | G4/50: 85% @ 4 Microns | 208EA Carbon Cell | Low volume vapours, Solvents, Conformal Coating, Adhesives, Airbrushing, Cleaning Fluids. NOT for MEK |
J – High Volume Fume | G4/50: 85% @ 4 Microns | 208EA Deep Bed Carbon Cell | Higher volume vapours, Solvents, Conformal Coating, Adhesives, Airbrushing, Cleaning Fluids. NOT for MEK |
R – Special Duty Fume | G4/25: Media Pad | Chemisorb Carbon Cell | Formaldehyde, Nitrogen Dioxides, Sulphur Dioxides, Hydrogen Sulphide, MEK’s |